Wir benötigen Ihre Einwilligung zur Verwendung der einzelnen Daten, damit Sie unter anderem Informationen zu Ihren Interessen einsehen können. Klicken Sie auf "OK", um Ihre Zustimmung zu erteilen.
Standard Test Method for Sulfide Ion in Water
Automatische name übersetzung:
Standardtestverfahren für Sulfide Ion in Wasser
NORM herausgegeben am 15.5.2009
Bezeichnung normen: ASTM D4658-09
Anmerkung: UNGÜLTIG
Ausgabedatum normen: 15.5.2009
SKU: NS-28061
Zahl der Seiten: 5
Gewicht ca.: 15 g (0.03 Pfund)
Land: Amerikanische technische Norm
Kategorie: Technische Normen ASTM
Keywords:
electrode, potentiometric, sulfide, water, Inorganic compounds--water, Ion selective electrode method, Sulfide content, ICS Number Code 13.060.50 (Examination of water for chemical substances)
Significance and Use | ||||||||||||||
Sulfide ion is found in ground waters and wastewater, causing odor and corrosion problems. If acidified, these waters can release hydrogen sulfide, which is extremely toxic even at low levels. This test method provides a means for interference-free measurement of free sulfide ion. Note 1—Sulfide forms complexes with hydrogen ions (HS1− and H2S). In addition, sulfide ion forms soluble complexes with elemental sulfur (S22−, S32−, S42−, etc.), tin, antimony, and arsenic ions. |
||||||||||||||
1. Scope | ||||||||||||||
1.1 This test method uses an ion-selective electrode to determine sulfide ion in water. The test method is applicable in the range from 0.04 to 4000 mg/L of sulfide. 1.2 Precision data presented in this test method were obtained using reagent water only. It is the user's responsibility to ensure the validity of this test method for untested types of water. 1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Sulfide samples, when acidified, can release highly toxic hydrogen sulfide gas. For a specific precautionary statement, see Note 2. |
||||||||||||||
2. Referenced Documents | ||||||||||||||
|
Wollen Sie sich sicher sein, dass Sie nur die gültigen technischen Normen verwenden?
Wir bieten Ihnen eine Lösung, die Ihnen eine Monatsübersicht über die Aktualität der von Ihnen angewandten Normen sicher stellt.
Brauchen Sie mehr Informationen? Sehen Sie sich diese Seite an.
Letzte Aktualisierung: 2024-11-04 (Zahl der Positionen: 2 209 323)
© Copyright 2024 NORMSERVIS s.r.o.