Standard Test Method for On-Line pH Measurement of Water of Low Conductivity
Automatische name übersetzung:
Standard Test Method for On-Line pH-Messung geringer Leitfähigkeit des Wassers
NORM herausgegeben am 1.1.2014
Bezeichnung normen: ASTM D5128-14
Anmerkung: UNGÜLTIG
Ausgabedatum normen: 1.1.2014
SKU: NS-29937
Zahl der Seiten: 11
Gewicht ca.: 33 g (0.07 Pfund)
Land: Amerikanische technische Norm
Kategorie: Technische Normen ASTM
Keywords:
high purity, low conductivity, on-line, pH, calculated pH, ICS Number Code 71.040.50 (Physicochemical methods of analysis)
Significance and Use | |||
5.1 Continuous pH measurements in low conductivity samples are sometimes required in pure water treatment using multiple pass reverse osmosis processes. Membrane rejection efficiency for several contaminants depends on pH measurement and control between passes where the conductivity is low. 5.2 Continuous pH measurement is used to monitor power plant cycle chemistry where small amounts of ammonia or amines or both are added to minimize corrosion by high temperature pure water and steam. 5.3 Conventional pH measurements are made in solutions that contain relatively large amounts of acid, base, or dissolved salts. Under these conditions, pH determinations may be made quickly and precisely. Continuous on-line pH measurements in water samples of low conductivity are more difficult 1.1 This test method covers the precise on-line determination of pH in water samples of conductivity lower than 100 μS/cm (see Table 1 and Table 2) over the pH range of 3 to 11 (see Sample |
Sample |
Sample |
Δ pH Error from Additional 1 mg/L NaOH Contaminate |
0.001 |
0.055 |
7.05 |
Δ 2.35 |
0.010 |
0.082 |
7.45 |
Δ 1.95 |
0.100 |
0.625 |
8.40 |
Δ 1.03 |
1.0 |
6.229 |
9.40 |
Δ 0.30 |
8.0 |
49.830 |
10.30 |
Δ 0.05 |
Sample
Concentration,
mg/L
Sample
Conductivity,
μS/cm
Sample
pH
Δ pH Error from Additional 1 mg/L HCl Contaminate
0.001
0.060
6.94
Δ2.38
0.010
0.134
6.51
Δ 1.95
0.100
1.166
5.56
Δ 1.03
1.0
11.645
4.56
Δ 0.30
8.0
93.163
3.66
Δ 0.05
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Standard Specification for Reagent Water |
|
Standard Terminology Relating to Water (Includes all amendments and changes 2/14/2024). |
|
Standard Test Method for pH Measurement of Water of Low Conductivity |
|
Standard Practice for Handling of High Purity Water Samples |
|
Standard Guide for On-Line Monitoring Systems for Water Analysis |
|
Standard Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water |
|
Standard Test Methods for pH of Water |
Historisch
1.11.2011
Historisch
1.1.2013
Historisch
1.11.2011
Historisch
1.4.2007
Historisch
1.10.2008
Historisch
1.1.2013
Wollen Sie sich sicher sein, dass Sie nur die gültigen technischen Vorschriften verwenden?
Wir bieten Ihnen Lösungen, damit Sie immer nur die gültigen (aktuellen) legislativen Vorschriften verwenden könnten.
Brauchen Sie mehr Informationen? Sehen Sie sich diese Seite an.
Letzte Aktualisierung: 2024-11-22 (Zahl der Positionen: 2 206 568)
© Copyright 2024 NORMSERVIS s.r.o.