Standard Test Method for On-Line Measurement of Low-Level Dissolved Oxygen in Water
Automatische name übersetzung:
Standard Test Method for On-Line Messung der Low-Level gelösten Sauerstoff im Wasser
NORM herausgegeben am 15.6.2013
Bezeichnung normen: ASTM D5462-13
Anmerkung: UNGÜLTIG
Ausgabedatum normen: 15.6.2013
SKU: NS-31326
Zahl der Seiten: 6
Gewicht ca.: 18 g (0.04 Pfund)
Land: Amerikanische technische Norm
Kategorie: Technische Normen ASTM
Keywords:
dissolved oxygen, galvanic cell, oxygen, polarographic sensor, ICS Number Code 13.060.50 (Examination of water for chemical substances)
Significance and Use | ||||||||||||
5.1 DO may be either a corrosive or passivating agent in boiler/steam cycles and is therefore controlled to specific concentrations that are low relative to environmental and wastewater treatment samples. Out-of-specification DO concentrations may cause corrosion in boiler systems, which leads to corrosion fatigue and corrosion products—all detrimental to the life and efficient operation of a steam generator. The efficiency of DO removal from boiler feedwater by mechanical or chemical means, or both, may be monitored by continuously measuring the DO concentration before and after the removal process with on-line instrumentation. DO measurement is also a check for air leakage into the boiler water cycle. 5.2 Guidelines for feedwater to high-pressure boilers with all volatile treatment generally require a feedwater DO concentration below 5 μg/L (3). 5.3 Boiler feedwater with oxygenated treatment is maintained in a range of 50 to 300 μg/L DO 5.4 In microelectronics production, DO can be detrimental in some manufacturing processes, for example, causing undesirable oxidation on silicon wafers. |
||||||||||||
1. Scope | ||||||||||||
1.1 This test method covers the on-line determination of dissolved oxygen (DO) in water samples primarily in ranges from 0 to 500 μg/L (ppb), although higher ranges may be used for calibration. On-line instrumentation is used for continuous measurements of DO in samples that are brought through sample lines and conditioned from high-temperature and high-pressure sources when necessary. 1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.3 This standard does not
purport to address all of the safety concerns, if any, associated
with its use. It is the responsibility of the user of this standard
to establish appropriate safety and health practices and determine
the applicability of regulatory limitations prior to use.
Standard Terminology Relating to Water
(Includes all amendments and changes 2/14/2024). Standard Specification for Reagent
Water Standard Practice for Sampling Steam
(Includes all amendments and changes 12/31/2018). Standard Guide for On-Line Monitoring
Systems for Water Analysis Standard Practices for Sampling Water
from Flowing Process Streams Standard Practice for Determination of
Precision and Bias of Applicable Test Methods of Committee D19 on
Water |
Historisch
1.5.2009
Historisch
1.6.2013
Historisch
15.6.2013
Historisch
1.5.2013
Historisch
15.11.2008
Historisch
15.6.2012
Wollen Sie sich sicher sein, dass Sie nur die gültigen technischen Normen verwenden?
Wir bieten Ihnen eine Lösung, die Ihnen eine Monatsübersicht über die Aktualität der von Ihnen angewandten Normen sicher stellt.
Brauchen Sie mehr Informationen? Sehen Sie sich diese Seite an.
Letzte Aktualisierung: 2025-01-16 (Zahl der Positionen: 2 219 847)
© Copyright 2025 NORMSERVIS s.r.o.