Standard Practice for Evaluation of Modular Connection of Proximally Fixed Femoral Hip Prosthesis
Automatische name übersetzung:
Standard Praxis für die Bewertung von Modular Anschluss von Proximal Feste Oberschenkel Hüftprothese
NORM herausgegeben am 1.2.2013
Bezeichnung normen: ASTM F2580-13
Anmerkung: UNGÜLTIG
Ausgabedatum normen: 1.2.2013
SKU: NS-54178
Zahl der Seiten: 5
Gewicht ca.: 15 g (0.03 Pfund)
Land: Amerikanische technische Norm
Kategorie: Technische Normen ASTM
Keywords:
arthroplasty, femoral hip prostheses, orthopedic medical devices, proximal fixation, total hip arthroplasty, ICS Number Code 11.040.40 (Implants for surgery, prothetics and orthotics)
Significance and Use | ||||||||
4.1 This practice can be used to describe the effects of materials, manufacturing, and design variables on the fatigue performance of metallic femoral hip prostheses subject to cyclic loading for large numbers of cycles. 4.2 The loading of femoral hip designs 4.3 In order for fatigue data on femoral hip prostheses to be comparable, reproducible, and capable of being correlated among laboratories, it is essential that uniform procedures be established. |
||||||||
1. Scope | ||||||||
1.1 This practice covers a procedure for the fatigue testing of metallic femoral hip prostheses used in hip joint replacements. This practice covers the procedures for the performance of fatigue tests on metallic femoral hip stems using a cyclic, constant-amplitude force. It applies to hip prostheses that utilize proximal metaphyseal fixation and are of a modular construct, and it is intended to evaluate the fatigue performance of the modular connections in the metaphyseal filling (that is, proximal body) region of the stem. 1.2 This practice is intended to provide useful, consistent, and reproducible information about the fatigue performance of metallic hip prostheses while held in a proximally fixated manner, with the distal end not held by a potting medium. 1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. |
||||||||
2. Referenced Documents | ||||||||
|